Detection Principle
This approach uses Pseudomonas sp. glutaminase to convert glutamine to glutamate + NH₄⁺. Two ISEs measure the differential signal: Electrode 1 (reference) measures baseline blood NH₄⁺ + K⁺ interference, while Electrode 2 (active) has immobilized glutaminase and measures baseline + enzyme-generated NH₄⁺. The differential signal = glutamine-derived ammonium only.

1. Preparation of Ion-Selective Membranes

1.1 NH₄⁺ Ion-Selective Membrane

  1. Prepare membrane cocktail: In a clean glass vial, mix the following components:
    • 1 mg Ammonium ionophore I (nonactin)
    • 33 mg PVC (high molecular weight)
    • 66 mg Plasticizer (DOS or o-NPOE)
    • 1 mL THF (tetrahydrofuran)
  2. Dissolve components: Stir vigorously for 2-3 hours until completely dissolved. The solution should be clear and homogeneous.
  3. Store properly: Seal the vial and store at room temperature. Use within 1 week.
Tip
Prepare the membrane cocktail in a fume hood due to THF vapors. Ensure all glassware is dry as water can cause PVC precipitation.

1.2 K⁺ Reference Membrane

  1. Prepare reference membrane cocktail:
    • 1 mg Valinomycin (K⁺ ionophore)
    • 33 mg PVC
    • 66 mg Plasticizer (dibutyl phthalate)
    • 1 mL THF
  2. Dissolve and store: Same procedure as NH₄⁺ membrane.

2. Electrode Body Preparation

2.1 Construction of ISE Electrodes

  1. Prepare electrode bodies: Use glass or plastic tubes (2 mm diameter, 5-10 cm length) as electrode bodies.
  2. Insert Ag/AgCl wire: Place a silver wire (0.5 mm diameter) inside each electrode body.
  3. Prepare internal solution: Fill the electrode body with 0.1 M KCl solution.
  4. Seal the reference: Ensure the Ag/AgCl wire is immersed in the internal solution and seal the bottom.

2.2 Membrane Application

  1. Prepare electrode tip: The tip of each electrode (where the membrane will be applied) should be polished flat and clean.
  2. Apply NH₄⁺ membrane: Using a micropipette, deposit 5 μL of NH₄⁺ membrane cocktail onto the electrode tip. Allow to dry in a desiccator for 2-3 hours.
  3. Apply additional layers: Repeat the application 2-3 times to achieve a membrane thickness of approximately 100-200 μm.
  4. Prepare two identical electrodes: Create two NH₄⁺ ISEs - one will serve as reference (Electrode 1) and one as active (Electrode 2).

3. Dual Electrode Assembly

  1. Mount electrodes: Secure both ISEs in a parallel configuration with a fixed spacing of 2-3 mm in an electrode holder.
  2. Connect to meter: Connect each electrode to a high-impedance voltmeter or pH meter capable of measuring potential difference.
  3. Label electrodes:
    • Electrode 1: Reference (no enzyme) - measures baseline NH₄⁺
    • Electrode 2: Active (with glutaminase) - measures baseline + glutamine-derived NH₄⁺
Key Point
The differential measurement approach is critical: both electrodes see identical [K⁺] (~4-5 mM in blood) and baseline NH₄⁺ (~20-50 μM). The difference (Electrode 2 - Electrode 1) represents only the glutamine-derived ammonium.

4. Quality Checks

  1. Potential stability test: Place both electrodes in 0.1 M PBS (pH 7.4) and monitor the potential for 5 minutes. Stable potentials (< 2 mV drift) indicate proper membrane formation.
  2. Nernstian response test: Test the electrodes in NH₄Cl standards (1-100 μM). A slope of 50-60 mV/decade indicates proper NH₄⁺ selectivity.
  3. Interference check: Verify K⁺ interference by comparing responses in solutions with and without 5 mM KCl.

5. Next Steps

After fabricating the dual ISE system, proceed to the Enzyme Immobilization Protocol to apply the Pseudomonas sp. glutaminase to the active electrode.