pH Compatibility Note
This approach uses Pseudomonas sp. glutaminase which maintains optimal activity at pH 7.0-7.5, making it fully compatible with whole blood pH (~7.4). The ammonium ISE method measures NH₄⁺ released when glutaminase converts glutamine to glutamate.

1. Ion-Selective Electrode Components

NH₄⁺ Ion-Selective Membrane
  • Ammonium ionophore I Nonactin or similar NH₄⁺ selective carrier
  • PVC matrix High molecular weight PVC for membrane
  • Plasticizer DOS (dioctyl sebacate) or o-NPOE
  • Tetrahydrofuran (THF) Anhydrous, for membrane casting
K⁺ Reference Ionophore
  • Potassium ionophore I Valinomycin or similar K⁺ carrier
  • PVC matrix For reference electrode membrane
  • Plasticizer DOS or dibutyl phthalate

2. Electrode Hardware

Electrode Bodies
  • Glass or plastic electrode bodies 2 mm diameter, 5-10 cm length
  • Internal reference electrode Silver/silver chloride wire
  • Internal filling solution 0.1 M KCl for Ag/AgCl
Connection Hardware
  • Coaxial cable Shielded, for low-noise measurement
  • BNC connectors For potentiometer connection
  • Electrode holder For stable mounting during measurement

3. Enzyme and Reagents

Enzyme
  • Glutaminase (Pseudomonas sp.) Recombinant, ≥40 U/mg, optimal pH 7.0-7.5
  • Recommended suppliers Sigma-Aldrich, Megazyme, Creative Enzymes
  • Quantity 100-250 units
Cross-linking
  • Bovine Serum Albumin (BSA) ≥96% purity, lyophilized
  • Glutaraldehyde 25% solution in water

4. Membrane and Support Materials

Enzyme Membrane Support
  • Porous cellulose acetate 0.2-0.45 μm pore size
  • Nylon mesh 20-50 μm pore size, for support
  • Porous silica membrane Alternative support material
Protective Membranes
  • Nafion solution 5% w/w in alcohol/water
  • Polyurethane Medical-grade, aliphatic polyether

5. Buffers and Solutions

Buffer Solutions
  • Phosphate Buffered Saline (PBS) 0.1 M, pH 7.4, for measurements
  • Phosphate buffer 50 mM, pH 7.4, for enzyme solution
  • KCl electrolyte 0.1 M, for internal filling
Standards
  • Ammonium chloride standard 1 M stock solution
  • L-Glutamine ≥99% purity
  • L-Glutamic acid ≥99% purity
  • Potassium chloride For interference testing

6. Equipment

Measurement
  • High-impedance voltmeter/pH meter ≥10¹² Ω input impedance
  • Data acquisition system For potential vs. time measurement
  • Temperature controller 25°C ± 0.5°C
Fabrication
  • Analytical balance 0.1 mg readability
  • Micropipettes 0.5-10 μL, 10-100 μL, 100-1000 μL
  • pH meter ±0.01 pH accuracy
  • Fume hood For THF handling
  • Vortex mixer For solution preparation

7. Key Differences from Approach 1

Component Approach 1 (Glutamate Oxidase) Approach 2 (Ammonium ISE)
Enzymes Glutaminase + Glutamate oxidase Glutaminase only
Electrodes Dual working electrodes (amperometric) Dual ISEs (potentiometric)
Signal Current (H₂O₂ oxidation) Potential (NH₄⁺ concentration)
Equipment Potentiostat High-impedance voltmeter